HHPC-6 Specifications
|
 |
| 6
Size Channels |
0.3
µm; 0.5 µm; 0.7 µm; 1.0 µm;
2.0 µm; 5.0 µm |
| Flow
Rate |
0.1
CFM (2.83 l/min.) |
| Light
Source |
Laser
Diode; Index Guided |
| MBTF |
>
36,000 hours @ 25¡ÆC |
| Calibration |
PSL
particles in air (NIST Traceable) |
| Counting
Efficiency |
50%
@ 0.3 µm; 100% for particles > 0.45 µm
(per JIS B9921) |
| Zero
Count |
<
1 count/5 minute (per JIS B9921) |
| Coincidence
Loss |
5%
at 2,000,000 particles per ft3 |
| Data
Storage |
500
Sample Records |
| Data
Recorded |
Date,
time, counts, R/H, temperature, sample volumes,
alarms, label |
| Display |
Graphic
liquid crystal with back light |
| Alarms |
Counts,
low battery, sensor fail |
| Count
Modes |
Concentration,
Totalize, Audio |
| Delay
Time |
0
~ 24 hours |
| Sample
Inlet |
Isokinetic
probe |
| Interface |
RS-232
and RS-485 via RJ-45 |
| Vacuum
Source |
Internal
pump flow controlled |
| Housing |
Injection
molded plastic with internal, conductive coating |
| Dimensions |
4.5"
(w) x 8.25" (l) x 2.25" (d); (11.43 cm x
20.96 cm x 5.72 cm) |
| Weight |
2.2
pounds, 1 kg |
|
|
| Power |
|
 |
| AC Adapter |
12 VDC at
2.5 A, 90 - 250 VAC, 50 -60 Hz |
| Rechargeable
Battery |
NiMH, 4.8
V at 4.5 Ah; Replaceable |
| Continuous
Operating Time |
8 hours |
| Recharge
Time |
2 hours |
|
|
|
Operating
Conditions
|
|
 |
| Relative
Humidity |
20
to 90%, non-condensing |
| Temperature |
10
to 40¡ÆC (50 to 104¡Æ F) |
|
|
|
Storage Conditions
|
|
 |
| Relative
Humidity |
Up
to 90%, non-condensing |
| Temperature |
-10
to 50¡ÆC (14 to 122¡ÆF) |
|
|
| Accessories |
Options |
 |
| Screw-on
Iso Probe |
Serial
Printer with Cable |
| Operation
Manual |
Carrying
Case |
| Utility
Software |
Zero Count
Filter |
| 1/8in.
Hose Barb Adapter |
|
|
|
 |
| Standards |
JIS B9921
|