Description
ATI manufactures portable Laskin nozzle aerosol generators which produce
a sub-micron polydispersed oil mist aerosol in concentrations from 10 to
100 micrograms per liter (ug/l) at air flows from 50 to 8,100 cfm @ 20
psig.
Aerosol generators and photometers are used to integrity test or locate
leaks in high efficiency air filtration systems. Filter manufacturers
use this equipment to scan ULPA and HEPA filters to verify they are free
from manufacturing defects. Filter certifiers use this equipment to
insure that filters were not damaged in shipping and have been installed
properly, eliminating any leakage.
With the proper generator and photometer combination, filter
deficiencies such as pinholes, thin spots, gasket leaks, frame leaks or
seal problems can be quickly and quantifiably pinpointed and corrected
thus protecting product and personnel.
Both the TDA-4B and the TDA-4B Lite feature several improvements
including rugged stainless steel construction, large fill port,
individual nozzle control, and a 3" standard sanitary fitting
outlet. An optional hose adapter, part # 9300100, is available for
introducing the aerosol into positive pressure systems.
The TDA-4B Lite is a smaller, lower capacity version of the TDA-4B that
still retains all the improved features. The TDA-4B lite was
specifically designed for use in biosafety cabinets. Its small size and
low cost make it the ideal generator for biosafety cabinet testing and
HEPA filter vacuums.
The TDA-4B Lite has 3 Laskin nozzles. When its total output at 20 psig
is diluted by 405 cfm of air, the aerosol concentration is approximately
100 ug/l. Two valves permit the unit to operate with 1 to 3 nozzles to
provide a wide range of aerosol concentrations.
TDA-4B Lite Specifications
Aerosol Output Range: 50-5,000 cfm
Aerosol Concentration 100 ug/1@*: 405 cfm
Aerosol Concentration 10 ug/1@*: 4,050 cfm
Generator Type: 1 to 3 Laskin Nozzles
Compressed Air: 3 to 9 cfm @ 20 psi
Aerosol Type: Polydispersed (Cold)
Size: 10" L Ąż 8" W Ąż 9" H
(25cm L Ąż 20cm W Ąż 23cm H)
Weight-Pounds (lbs): 12 lbs.
Weight-Kilograms (kg): 5.5 kg
Electrical: (Not Required)
*When operated at 20 psi.
|