Radiography Survey Meter |
Model 3009A |
The Model 3009A survey meter is a specialized instrument designed for industrial radiography. It meets the requirements of Title10, Part 34 of the Code of Federal Regulations (requirements for radiographers). It features an energy compensated, internal detector capable of accurately detecting a broad range of exposure rates and photon energies. A top-mounted, splash-proof speaker is provided, as well as separate calibration controls for each range. FEATURES: • Speaker for easier more reliable surveys • Internal detector design protects GM tube from rough handling and results in less down time. • Separate calibration controls for each range • Meets 10CFR34 requirements |
|
| OPERATING CHARACTERISTICS | 2 | |||||
| Radiation Detected: | Gamma and X-rays | |||||
| Accuracy: | ‐20% | |||||
| Energy Dependency: | ‐30% 80 keV to 1.25 Mev | |||||
| Linearity: | ‐20% | |||||
| Operating Ranges: | 0 - 10 mR/h; 0 - 100 mR/h; 0 - 1000 mR/h (0.1 mSv/h; 0-1 mSv/h) | |||||
| Exposure Rate Limit: | 100 R/h (1 Sv/h) | |||||
| Warm Up Time: | None | |||||
| Response Time: | 5 second time constant | |||||
| Environmental Effects: |
Temperature Range: 14━F to 122━F (-10━C to
50━C) Temperature Dependence: 15% throughout operating range Relative Humidity Range: 20% to 95% (non-condensing) |
|||||
| PHYSICAL CHARACTERISTICS | 2 | |||||
| Detector Type: | Energy Compensated Halogen Quenched GM Tube (Model 1-7301) | |||||
| Detector Voltage: | 920V | |||||
| Readout: Visual Audio |
2.62" round meter Speaker, clicking type headset (optional) |
|||||
| Controls: | 4 position selector switch (off, x1, x10, x100) Battery check switch, Speaker on-off switch | |||||
| Power: Batteries |
One 9 volt cell for speaker (NEDA 1604-A or equal) 2 - "D" cells (1.5 volt) |
|||||
| Connectors: | Phonejack for Model 3020 headset | |||||
| Mechanical: |
|
|||||
| ACCESSORIES | 2 | |||||
| Included: | Mounting: Carrying Strap | |||||
| Controls: | Audio: Model 3020 Headset |