QMI.JPG (6876 bytes)   QMI Inc.    AIRSCAN   INSTRUMENTS   SCANNERS    ULTRASONIC TESTING    REQUESTS    ADDRESSES

SONDA 007CX

Multifrequency Instrument For Air-Coupled Ultrasonic Testing

 The SONDA 007CX AIRSCAN is an air-coupled ultrasonic flaw detector. The ambient air is used as the couplant. It drives very powerfull transducers with tonebursts (selectable at 50 kHz, 200 kHz, 400 kHz, and 1 MHz). The receiver includes a narrow band super low-noise and high-gain preamplifier.

Sonda 007 CX

SONDA 007CX AIRSCAN

 

The SONDA can be assembled and mounted in a cabinet as a MULTICHANNEL AIRSCAN Instrument

Two-Channel SONDA

 

FEATURES

Easy Storage and Retrieval of Test Setup Parameters
Stand-Alone and Remote Operation via Host Computer
Outlet for C-Scan Imaging
Self-Contained Function Control

airres_s.TIF (30386 bytes)

400 kHz

Using 400 kHz AS 400 air-coupled focused transducers, a lateral resolution of about 1.5 mm (0.060") is achieved, as demonstrated by this C-scan with artificial defects

50kHz on Foam

50 kHz

With 50 kHz transducers it is possible to penetrate very attenuative materials, such as foam, foam sandwiches and honeycomb structures, and to obtain C-scan images.

One-sided Bubbler System

120 kHz / 200 kHz

New 120 kHz air-coupled transducers are used for  pultruded materials, yielding an effective compromise between resolution and penetration.

The SONDA 007CX can also be combined with contact methods, such as water squirters  (as shown in the picture in a pitch-catch configuration) and roller probes, to test parts which have very high attenuation.

 

ACCESSORIES

Air Coupled Ultrasonic Transducers for 50, 120, and 400 kHz
Preamplifier LNPA 600
Dry Contact Roller Probes
Soft Tip Probes
200 kHz Squirter Transducer Assemblies

CONFIGURATIONS

The SONDA-007CX can be used with all QMI scanning systems and can be retrofitted to most customers' systems.

 

SPECIFICATIONS

Transmitter Bipolar toneburst - up to 15 cycles per burst.
Fine tuning for nominal 50 kHz, 200 kHz, 400 kHz and 1 MHz center frequencies.
Pulse repetition rate adjustable up to 500 Hz.
Receiver Fine tuning for all frequencies.
Remote Preamplifier LNPA 501A add-on super low noise preamplifier (600pV/root Hz).
Gain 30 dB, output impedance 50 ohm (can drive cables over 100 feet long).
Preamplifier built in the 50, 200, and 400 kHz receiving transducer.
Main Receiver Amplifier Course Attenuator: 0, -20, -40, -60 dB
Fixed Gain: + 40 Db
Variable Gain: 0 to 40 dB
CRT Display Flat Panel Digital 7 in. diagonal display.
A-trace RF and FW rectified envelope including menu function display.
Front Keypad Sealed dedicated function control push buttons.
Rear Panel DC analog output. RS 232 connector for host computer.
Video connector for additional screen monitor.
Power Requirement 110 - 120 Volt, 5 Amp.
220 - 230 Volt, 2.5 Amp.
Operating Requirement -28 deg C to 38 deg C
-20 deg F to 100 deg F

 